Objective: To design an improved accuracy filter for photoplethysmography Direct Current (DC) component denoising and clinically applicable measurement design.
Methodology: The data source for this study was obtained from Physionet. The obtained PPG signal was mixed with seven different types of noises. The seven different derivates of PPG signals were denoised using 4 types of denoisers that were designed using Infinite impulse response (IIR), finite impulse response (FIR), Kalman, and improved Kalman filters. The difference can be measured by the SNR value, the lower signal to noise ratio (SNR) value shows higher noise in the signal.
Results: SNR value achieved on FIR filter was 7.54dB, IIR 9.08dB, Kalman 18.96dB and the maximum SNR achieved on the improved Kalman filter which is 24.76 dB.
Conclusion: It can be said that the Improved Kalman Filter is more appropriate for DC signal denoising in PPG signals rather than the FIR filter, IIR filter, and Kalman filter.

Key words: PPG signal, motion artifact, Kalman filter, Health care.


 
ARTICLE TOOLS
Abstract
PDF Fulltext
How to cite this articleHow to cite this article
Citation Tools
Related Records
 Articles by Muhammad Kashif
Articles by Muhammad Fahmi
Articles by Dita Aprilia Hariyani
Articles by Akif Rahmatilah
Articles by Khusnul Ain
Articles by Yunus Susilo
Articles by Ardiyansyah Syahrom
Articles by Suryani Dyah Astuti
on Google
on Google Scholar


How to Cite this Article
Pubmed Style

Kashif M, Fahmi M, Hariyani DA, Rahmatilah A, Ain K, Susilo Y, Syahrom A, Astuti SD. An improved Kalman Filter in Photoplethysmography DC Component Denoising for cardiorespiratory analysis. RMJ. 2022; 47(3): 741-745. doi:10.5456/rmj.2004


Web Style

Kashif M, Fahmi M, Hariyani DA, Rahmatilah A, Ain K, Susilo Y, Syahrom A, Astuti SD. An improved Kalman Filter in Photoplethysmography DC Component Denoising for cardiorespiratory analysis. https://www.rmj.org.pk/?mno=48464 [Access: December 04, 2023]. doi:10.5456/rmj.2004


AMA (American Medical Association) Style

Kashif M, Fahmi M, Hariyani DA, Rahmatilah A, Ain K, Susilo Y, Syahrom A, Astuti SD. An improved Kalman Filter in Photoplethysmography DC Component Denoising for cardiorespiratory analysis. RMJ. 2022; 47(3): 741-745. doi:10.5456/rmj.2004



Vancouver/ICMJE Style

Kashif M, Fahmi M, Hariyani DA, Rahmatilah A, Ain K, Susilo Y, Syahrom A, Astuti SD. An improved Kalman Filter in Photoplethysmography DC Component Denoising for cardiorespiratory analysis. RMJ. (2022), [cited December 04, 2023]; 47(3): 741-745. doi:10.5456/rmj.2004



Harvard Style

Kashif, M., Fahmi, . M., Hariyani, . D. A., Rahmatilah, . A., Ain, . K., Susilo, . Y., Syahrom, . A. & Astuti, . S. D. (2022) An improved Kalman Filter in Photoplethysmography DC Component Denoising for cardiorespiratory analysis. RMJ, 47 (3), 741-745. doi:10.5456/rmj.2004



Turabian Style

Kashif, Muhammad, Muhammad Fahmi, Dita Aprilia Hariyani, Akif Rahmatilah, Khusnul Ain, Yunus Susilo, Ardiyansyah Syahrom, and Suryani Dyah Astuti. 2022. An improved Kalman Filter in Photoplethysmography DC Component Denoising for cardiorespiratory analysis. Rawal Medical Journal, 47 (3), 741-745. doi:10.5456/rmj.2004



Chicago Style

Kashif, Muhammad, Muhammad Fahmi, Dita Aprilia Hariyani, Akif Rahmatilah, Khusnul Ain, Yunus Susilo, Ardiyansyah Syahrom, and Suryani Dyah Astuti. "An improved Kalman Filter in Photoplethysmography DC Component Denoising for cardiorespiratory analysis." Rawal Medical Journal 47 (2022), 741-745. doi:10.5456/rmj.2004



MLA (The Modern Language Association) Style

Kashif, Muhammad, Muhammad Fahmi, Dita Aprilia Hariyani, Akif Rahmatilah, Khusnul Ain, Yunus Susilo, Ardiyansyah Syahrom, and Suryani Dyah Astuti. "An improved Kalman Filter in Photoplethysmography DC Component Denoising for cardiorespiratory analysis." Rawal Medical Journal 47.3 (2022), 741-745. Print. doi:10.5456/rmj.2004



APA (American Psychological Association) Style

Kashif, M., Fahmi, . M., Hariyani, . D. A., Rahmatilah, . A., Ain, . K., Susilo, . Y., Syahrom, . A. & Astuti, . S. D. (2022) An improved Kalman Filter in Photoplethysmography DC Component Denoising for cardiorespiratory analysis. Rawal Medical Journal, 47 (3), 741-745. doi:10.5456/rmj.2004