RMJ. 2013; 38(4): 354-357


Accuracy of Accutrend GCT meter for the measurement of blood cholesterol levels

Abdul Rehman Arshad.

Abstract
Objective: This study was done to assess the accuracy of Accutrend GCT meter in determining blood cholesterol levels in our setup, with an aim to recommend its use in clinical practice
Methods: This cross sectional study was conducted on 251 healthy volunteers from April to May 2013. Capillary whole blood samples were obtained using sterile lancets and cholesterol levels were measured using Accutrend GCT meter (Roche Diagnostics, Mannheim, Germany) using the standard technique. Three ml whole blood was also collected in plain tubes by the same operator without applying a tourniquet. Serum cholesterol levels were determined using Merck Microlab 300 Automated Clinical Chemistry Analyser the same day. The laboratory technician performing this test was blinded to results obtained with Accutrend GCT meter. The seven patients with readings beyond the analytic range of the Accutrend GCT meter were excluded, meaning that data analysis was performed on 244 results only.
Results: Results of 182 males and 62 females having a mean age of 38.54± 15.25 years were analysed. There was a significant difference in mean cholesterol levels measured by Accutrend GCT and Microlab-300 (4.73± 0.74mmol/l and 4.67± 0.86 mmol/l; p=0.021), producing a measurement bias of 1.43% with Accutrend GCT meter. A strong positive correlation was found between the results of the two techniques (r=0.852 ,p<0.001).
Conclusions: Accutrend GCT meter provides accurate measurement of blood cholesterol levels and its use should be encouraged for point-of-care measurements.

Key words: Cholesterol, Lipids, Bias, Point-of-Care Systems, Point-of-Care Technology


 
ARTICLE TOOLS
Abstract
PDF Fulltext
How to cite this articleHow to cite this article
Citation Tools
Related Records
 Articles by Abdul Rehman Arshad
on Google
on Google Scholar

How to Cite this Article
Pubmed Style

Abdul Rehman Arshad. Accuracy of Accutrend GCT meter for the measurement of blood cholesterol levels. RMJ. 2013; 38(4): 354-357.


Web Style

Abdul Rehman Arshad. Accuracy of Accutrend GCT meter for the measurement of blood cholesterol levels. https://www.rmj.org.pk/?mno=40594 [Access: December 05, 2023].


AMA (American Medical Association) Style

Abdul Rehman Arshad. Accuracy of Accutrend GCT meter for the measurement of blood cholesterol levels. RMJ. 2013; 38(4): 354-357.



Vancouver/ICMJE Style

Abdul Rehman Arshad. Accuracy of Accutrend GCT meter for the measurement of blood cholesterol levels. RMJ. (2013), [cited December 05, 2023]; 38(4): 354-357.



Harvard Style

Abdul Rehman Arshad (2013) Accuracy of Accutrend GCT meter for the measurement of blood cholesterol levels. RMJ, 38 (4), 354-357.



Turabian Style

Abdul Rehman Arshad. 2013. Accuracy of Accutrend GCT meter for the measurement of blood cholesterol levels. Rawal Medical Journal, 38 (4), 354-357.



Chicago Style

Abdul Rehman Arshad. "Accuracy of Accutrend GCT meter for the measurement of blood cholesterol levels." Rawal Medical Journal 38 (2013), 354-357.



MLA (The Modern Language Association) Style

Abdul Rehman Arshad. "Accuracy of Accutrend GCT meter for the measurement of blood cholesterol levels." Rawal Medical Journal 38.4 (2013), 354-357. Print.



APA (American Psychological Association) Style

Abdul Rehman Arshad (2013) Accuracy of Accutrend GCT meter for the measurement of blood cholesterol levels. Rawal Medical Journal, 38 (4), 354-357.