Are blood glucose monitors for home use and hospital use interchangeable?

Sana Ansari, Adnan Mustafa Zubairi, Mohammad Salman.

Abstract
Objective: To assess the analytical performance of Accu check Active®, One Touch Ultra®, On Call Plus® and Nova Stat Strip® glucose monitoring point of care devices, when compared to each other and with Roche ® Hexokinase method on Hitachi 902 ® automated chemistry analyzer in the presence of interferants like hematocrits and ascorbic acid.
Methodology: The study was conducted at Dr. Ziauddin Hospital, Karachi from May 2014 to September 2014. Venous blood samples were modified pre-analytically to generate varying hematocrits. Samples were glycolyzed or spiked to achieve various glucose levels. Similarly, different levels of ascorbic acid were also added and glucose tested on blood glucose monitoring devices and automated chemistry analyzer. This process generated 18 different combinations of glucose with hematocrit or ascorbic acid.
Results: Differences in hematocrit and ascorbic acid levels brought about significant changes in glucose measurements in the glucometers. The Nova Stat Strip® displayed results close to the analyzer in the presence of interfering substances. The Accucheck Active® and One Touch Ultra® had a similar performance, while the On Call Plus® showed the most variability of results.
ISO 15197:2013 requirements were fulfilled by Nova Stat Strip® and Accucheck Active® while the One Touch Ultra® and On Call Plus® did not reach the acceptable limits.
Conclusion: Hospitalized patients have a multitude of interfering substances in their blood, which bring about a significant difference in glucose readings. Glucose monitoring devices designed with additional strips to minimize interference by interfering agents are more accurate and should be preferred over other devices for use in hospitals point of care testing and self monitoring.

Key words: Glucose, blood glucose self monitoring, point of care systems.


 
ARTICLE TOOLS
Abstract
PDF Fulltext
Print this article Print this Article
How to cite this articleHow to cite this article
Export to
Export to
Related Records
 Articles by Sana Ansari
Articles by Adnan Mustafa Zubairi
Articles by Mohammad Salman
on Google
on Google Scholar
Article Statistics
 Viewed: 132
Downloaded: 15
Cited: 0


How to Cite this Article
Pubmed Style

Ansari S, Zubairi AM, Salman M. Are blood glucose monitors for home use and hospital use interchangeable?. RMJ. 2017; 42(3): 390-395.



Web Style

Ansari S, Zubairi AM, Salman M. Are blood glucose monitors for home use and hospital use interchangeable?. www.scopemed.org/?mno=227191 [Access: October 20, 2017].



AMA (American Medical Association) Style

Ansari S, Zubairi AM, Salman M. Are blood glucose monitors for home use and hospital use interchangeable?. RMJ. 2017; 42(3): 390-395.



Vancouver/ICMJE Style

Ansari S, Zubairi AM, Salman M. Are blood glucose monitors for home use and hospital use interchangeable?. RMJ. (2017), [cited October 20, 2017]; 42(3): 390-395.



Harvard Style

Ansari, S., Zubairi, A. M. & Salman, M. (2017) Are blood glucose monitors for home use and hospital use interchangeable?. RMJ, 42 (3), 390-395.



Turabian Style

Ansari, Sana, Adnan Mustafa Zubairi, and Mohammad Salman. 2017. Are blood glucose monitors for home use and hospital use interchangeable?. Rawal Medical Journal, 42 (3), 390-395.



Chicago Style

Ansari, Sana, Adnan Mustafa Zubairi, and Mohammad Salman. "Are blood glucose monitors for home use and hospital use interchangeable?." Rawal Medical Journal 42 (2017), 390-395.



MLA (The Modern Language Association) Style

Ansari, Sana, Adnan Mustafa Zubairi, and Mohammad Salman. "Are blood glucose monitors for home use and hospital use interchangeable?." Rawal Medical Journal 42.3 (2017), 390-395. Print.



APA (American Psychological Association) Style

Ansari, S., Zubairi, A. M. & Salman, M. (2017) Are blood glucose monitors for home use and hospital use interchangeable?. Rawal Medical Journal, 42 (3), 390-395.




Instructions for Authors

AUTHOR LOGIN

REVIEWER LOGIN

Indexed In

WHO Index Medicus IMEMR,
Emromedex,
Pakmedinet,
ExtraMED and
Scopus

Approved by the Higher Education Commission of Pakistan and Pakistan Medical and Dental Council



The articles in Rawal Medical Journal are open access articles licensed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-sa/3.0/) which permits unrestricted, non-commercial use, distribution and reproduction in any medium, provided the work is properly cited.